Merit Medical Systems, Inc. Recall 95330
Description: Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.
Merit Medical Systems, Inc. Recall 95330 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0062-2025 |
Event ID | 95330 |
Event Description | Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid. |
Product Type | Devices |
Distribution | Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Japan, Monaco, Netherlands, Qatar, United Arab Emirates, United Kingdom. |
Quantity | 432 devices |
Recall Reason | Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier. |
Device Classification | 20241023 |
Device Code Info | Catalog Number: K05T-02159 UDI-DI code: 00884450140840 Lot Number: T2417859 T2434208 T2509792 T2549931 T2561788 T2625814 T2675576 T2762716 T2801089 T2893877 |
Center Classification Date | 20241011 |
Recall Initiation Date | 20240830 |
Recalling Firm | Merit Medical Systems, Inc. |
Initial Notification | Letter |
Similar To |