Baxter Healthcare Corporation Recall 90848

Description: Allen Standard Starburst, Mayfield/Doro, STANDARD SPINE POSITIONERS CMPT, REF A-70710, #800103. Component to position and support the patients head in surgical procedures.

Baxter Healthcare Corporation Recall 90848 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0177-2023
Event ID90848
Event DescriptionAllen Standard Starburst, Mayfield/Doro, STANDARD SPINE POSITIONERS CMPT, REF A-70710, #800103. Component to position and support the patients head in surgical procedures.
Product TypeDevices
DistributionMI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
Quantity2 units
Recall Reasonthe adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
Device Classification20221109
Device Code InfoUDI/DI 00615521002725, Lot Number: 22-123
Center Classification Date20221103
Recall Initiation Date20220920
Recalling FirmBaxter Healthcare Corporation
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit