Baxter Healthcare Corporation Recall 95409
Description: Baxter Operating Table TruSystem 7000 (dV) V, Product code REF 1841083
Baxter Healthcare Corporation Recall 95409 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0216-2025 |
Event ID | 95409 |
Event Description | Baxter Operating Table TruSystem 7000 (dV) V, Product code REF 1841083 |
Product Type | Devices |
Distribution | Worldwide distribution. |
Quantity | N/A |
Recall Reason | The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke. |
Device Classification | 20241030 |
Device Code Info | UDI/DI 00887761974227, All Serial Numbers |
Center Classification Date | 20241024 |
Recall Initiation Date | 20240919 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Letter |
Similar To |