Product Safety Recalls

Product Recall Tracker

Siemens Healthcare Diagnostics, Inc. Recall 95751

Description: Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

Siemens Healthcare Diagnostics, Inc. Recall 95751 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0767-2025
Event ID95751
Event DescriptionMagnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain.
Quantity13,115 units (11,520 US, 1,595 ROW)
Recall ReasonSiemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patient s results are incorrect, and the results may be reported out
Device Classification20250101
Device Code InfoSiemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC.
Center Classification Date20241226
Recall Initiation Date20241112
Recalling FirmSiemens Healthcare Diagnostics, Inc.