Shimadzu Medical Systems Recall 89596

Description: MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

Shimadzu Medical Systems Recall 89596 Information

StatusOngoing
Mandated?FDA Mandated
Recall NumberZ-0869-2022
Event ID89596
Event DescriptionMODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Product TypeDevices
DistributionU.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
Quantity12 systems
Recall ReasonIt was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Device Classification20220413
Device Code InfoModel: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013
Center Classification Date20220406
Recall Initiation Date20220110
Recalling FirmShimadzu Medical Systems
Initial Notification Letter