SIN SISTEMA DE IMPLANTE NACIONAL S.A Recall 95983

Description: Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

SIN SISTEMA DE IMPLANTE NACIONAL S.A Recall 95983 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0920-2025
Event ID95983
Event DescriptionImplant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
Product TypeDevices
DistributionUS distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
Quantity123 (US 98, OUS 25)
Recall ReasonDental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Device Classification20250122
Device Code InfoUDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01
Center Classification Date20250115
Recall Initiation Date20240901
Recalling FirmSIN SISTEMA DE IMPLANTE NACIONAL S.A