Philips North America Llc Recall 88461
Description: Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781358
Philips North America Llc Recall 88461 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2344-2021 |
Event ID | 88461 |
Event Description | Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781358 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom. |
Quantity | 6 units; 12 OUS |
Recall Reason | Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury |
Device Classification | 20210901 |
Device Code Info | S/N and UDI: US: 45245 (01)00884838088115(21)45245; 45247 (01)00884838088115(21)45247; 45261 (01)00884838088115(21)45261; 45263 (01)00884838088115(21)45263; 45264 (01)00884838088115(21)45264; 45269 (01)00884838088115(21)45269; OUS: 45215 (01)00884838088115; 45284 (01)00884838088115; 45253 (01)00884838088115; 45313 (01)00884838088115; 45265 (01)00884838088115; 45297 (01)00884838088115; 45218 (01)00884838088115; 45272 (01)00884838088115; 45281 (01)00884838088115; 45231 (01)00884838088115; 45278 (01)00884838088115; 45279 (01)00884838088115; 45245 (01)00884838088115; 45247 (01)00884838088115; |
Center Classification Date | 20210825 |
Recall Initiation Date | 20210730 |
Recalling Firm | Philips North America Llc |
Similar To |