K2M, Inc Recall 88630
Description: Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
K2M, Inc Recall 88630 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0104-2022 |
Event ID | 88630 |
Event Description | Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068 |
Product Type | Devices |
Distribution | Domestic distribution to CO, IL, LA, MA, MS, OK, PA, TN, and TX. Foreign distribution to Columbia, UK, Netherlands, and Canada. |
Quantity | 23 |
Recall Reason | Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery. |
Device Classification | 20211020 |
Device Code Info | Lots CCUNA, HYMJ, KNCU |
Center Classification Date | 20211013 |
Recall Initiation Date | 20210908 |
Recalling Firm | K2M, Inc |
Initial Notification | Letter |