Philips Medical Systems Gmbh, DMC Recall 88393
Description: CombiDiagnost R90 Software Version R1.0 and R1.1
Philips Medical Systems Gmbh, DMC Recall 88393 Information
Status | Ongoing |
Mandated? | FDA Mandated |
Recall Number | Z-2357-2021 |
Event ID | 88393 |
Event Description | CombiDiagnost R90 Software Version R1.0 and R1.1 |
Product Type | Devices |
Distribution | US Nationwide Distribution |
Quantity | 322 systems; 66 in US |
Recall Reason | Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image. |
Device Classification | 20210908 |
Device Code Info | Model 706100 |
Center Classification Date | 20210828 |
Recall Initiation Date | 20200722 |
Recalling Firm | Philips Medical Systems Gmbh, DMC |
Initial Notification | Letter |
Similar To |