Product Safety Recalls

Product Recall Tracker

Elekta, Inc. Recall 88860

Description: Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

Elekta, Inc. Recall 88860 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0182-2022
Event ID88860
Event DescriptionOncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00
Product TypeDevices
DistributionDistribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam
Quantity1072 devices
Recall ReasonUnder certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.
Device Classification20211103
Device Code InfoOncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00 GTIN Numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700
Center Classification Date20211028
Recall Initiation Date20211013
Recalling FirmElekta, Inc.