Randox Laboratories Ltd. Recall 88130
Description: NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Randox Laboratories Ltd. Recall 88130 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2099-2021 |
Event ID | 88130 |
Event Description | NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115 |
Product Type | Devices |
Distribution | US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX |
Quantity | 71 kits US |
Recall Reason | Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples |
Device Classification | 20210728 |
Device Code Info | Batch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066 |
Center Classification Date | 20210720 |
Recall Initiation Date | 20210510 |
Recalling Firm | Randox Laboratories Ltd. |