Instrumentation Laboratory Recall 88930
Description: HemosIL ReadiPlasTin (20 mL), Part number 0020301400
Instrumentation Laboratory Recall 88930 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0276-2022 |
Event ID | 88930 |
Event Description | HemosIL ReadiPlasTin (20 mL), Part number 0020301400 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. |
Quantity | 5828 kits (310 US) |
Recall Reason | The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result. |
Device Classification | 20211201 |
Device Code Info | Lot N0705526 UDI 08426950632887 |
Center Classification Date | 20211123 |
Recall Initiation Date | 20211013 |
Recalling Firm | Instrumentation Laboratory |
Initial Notification | |
Similar To |