Siemens Healthcare Diagnostics, Inc. Recall 88086
Description: The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use.
Siemens Healthcare Diagnostics, Inc. Recall 88086 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2014-2021 |
Event ID | 88086 |
Event Description | The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam. |
Quantity | 842 units |
Recall Reason | ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result |
Device Classification | 20210714 |
Device Code Info | Software version: 1.4 or lower UDI - 00630414595467 |
Center Classification Date | 20210706 |
Recall Initiation Date | 20210518 |
Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
Initial Notification | Letter |