The Biomed Guys Recall 87768

Description: Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

The Biomed Guys Recall 87768 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1651-2021
Event ID87768
Event DescriptionAlaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.
Product TypeDevices
DistributionUs distribution to the states of IN and TX.
Quantity22 units
Recall ReasonBezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Device Classification20210609
Device Code InfoSerial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,
Center Classification Date20210601
Recall Initiation Date20210325
Recalling FirmThe Biomed Guys