Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Recall 87654
Description: CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Recall 87654 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1522-2021 |
Event ID | 87654 |
Event Description | CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED). |
Product Type | Devices |
Distribution | Worldwide |
Quantity | 16,311 devices |
Recall Reason | There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT). |
Device Classification | 20210512 |
Device Code Info | All serial numbers |
Center Classification Date | 20210430 |
Recall Initiation Date | 20210406 |
Recalling Firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) |
Initial Notification | Letter |
Similar To |