STEP-HAR MEDICAL LLC Recall 87773

Description: Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part

STEP-HAR MEDICAL LLC Recall 87773 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1543-2021
Event ID87773
Event DescriptionAlaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part
Product TypeDevices
DistributionIA
Quantity2
Recall ReasonBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Device Classification20210519
Device Code InfoSerial Numbers: 12747983 and 9919591
Center Classification Date20210510
Recall Initiation Date20210226
Recalling FirmSTEP-HAR MEDICAL LLC
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit