Product Safety Recalls

Product Recall Tracker

The Binding Site Group, Ltd. Recall 86926

Description: Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A

The Binding Site Group, Ltd. Recall 86926 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0943-2021
Event ID86926
Event DescriptionOptilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A
Product TypeDevices
DistributionU.S.: CO, FL, IA, MA, NC, NJ, NY, OH, SD, and UT. O.U.S.: Not provided
Quantity213 kits in the U.S.
Recall ReasonDue to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.
Device Classification20210210
Device Code InfoModel Number: LK016.M.OPT.A UDI: 05051700019866 Lot Number: 454763
Center Classification Date20210201
Recall Initiation Date20201118
Recalling FirmThe Binding Site Group, Ltd.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.