Product Safety Recalls

Product Recall Tracker

Roche Diagnostics Operations, Inc. Recall 86951

Description: CoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837738001

Roche Diagnostics Operations, Inc. Recall 86951 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0804-2021
Event ID86951
Event DescriptionCoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837738001
Product TypeDevices
DistributionUS Nationwide
Quantity716,880 units
Recall ReasonPossible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
Device Classification20210113
Device Code InfoAll serial numbers
Center Classification Date20210107
Recall Initiation Date20201202
Recalling FirmRoche Diagnostics Operations, Inc.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.