Roche Diagnostics Operations, Inc. Recall 86981

Description: Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001

Roche Diagnostics Operations, Inc. Recall 86981 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0755-2021
Event ID86981
Event DescriptionRoche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity7 units
Recall ReasonPotential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
Device Classification20210106
Device Code InfoSerial Numbers: 1925-10; 1931-04; 1931-05;1931-07;1937-10; 1938-01; 1942-10
Center Classification Date20201230
Recall Initiation Date20201208
Recalling FirmRoche Diagnostics Operations, Inc.