Intuitive Surgical, Inc. Recall 89109
Description: da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A
Intuitive Surgical, Inc. Recall 89109 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0419-2022 |
Event ID | 89109 |
Event Description | da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of Arizona, California, Connecticut, Florida, Illinois, Kansas, Kentucky, Massachusetts, Minnesota, Missouri, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington, and Wisconsin. |
Quantity | 46 |
Recall Reason | Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot. |
Device Classification | 20220105 |
Device Code Info | All da Vinci SP Surgical systems with software version OS4 v4.0.0 (P4_B327) |
Center Classification Date | 20211227 |
Recall Initiation Date | 20210922 |
Recalling Firm | Intuitive Surgical, Inc. |
Initial Notification | Letter |