XTANT Medical Recall 89367
Description: X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System.
XTANT Medical Recall 89367 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0533-2022 |
Event ID | 89367 |
Event Description | X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System. |
Product Type | Devices |
Distribution | U.S.: CA, MI, NV, and TX O.U.S.: Portugal |
Quantity | 24 devices |
Recall Reason | The proximal opening of the inserts exhibit an out of specification condition which would prevent the crossbar plates from appropriately engaging with the insert. |
Device Classification | 20220209 |
Device Code Info | X060-0270, 8mm Titanium Insert, Axle System UDI (Primary DI Number): M697X06002701/$$7058963W Lot Number: 058963 |
Center Classification Date | 20220128 |
Recall Initiation Date | 20211229 |
Recalling Firm | XTANT Medical |
Initial Notification | Letter |
Similar To |