Synthes (USA) Products LLC Recall 89428
Description: Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
Synthes (USA) Products LLC Recall 89428 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0634-2022 |
Event ID | 89428 |
Event Description | Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001 |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 600 units |
Recall Reason | Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection. |
Device Classification | 20220216 |
Device Code Info | Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372 |
Center Classification Date | 20220210 |
Recall Initiation Date | 20211222 |
Recalling Firm | Synthes (USA) Products LLC |
Initial Notification | Letter |
Similar To |