PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall 89437
Description: Zenition 50, Model #718096
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall 89437 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0647-2022 |
Event ID | 89437 |
Event Description | Zenition 50, Model #718096 |
Product Type | Devices |
Distribution | Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution. |
Quantity | 16 US; 76 ROW |
Recall Reason | Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system. |
Device Classification | 20220223 |
Device Code Info | Wireless Foot switch 3P : 459801238191 Wireless Foot switch 3P: 459801238231 |
Center Classification Date | 20220217 |
Recall Initiation Date | 20211202 |
Recalling Firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
Initial Notification | Letter |