Roche Molecular Systems, Inc. Recall 89643

Description: cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190

Roche Molecular Systems, Inc. Recall 89643 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0828-2022
Event ID89643
Event Descriptioncobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.
Quantity14 units
Recall ReasonRoche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
Device Classification20220406
Device Code InfoUDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086
Center Classification Date20220328
Recall Initiation Date20211116
Recalling FirmRoche Molecular Systems, Inc.