GE Healthcare, LLC Recall 89762
Description: Revolution Apex, Revolution CT with Apex Edition
GE Healthcare, LLC Recall 89762 Information
Status | Ongoing |
Mandated? | FDA Mandated |
Recall Number | Z-0815-2022 |
Event ID | 89762 |
Event Description | Revolution Apex, Revolution CT with Apex Edition |
Product Type | Devices |
Distribution | Worldwide - US Nationwide distribution. |
Quantity | 273 (51 US, 222 OUS) in total |
Recall Reason | The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high. |
Device Classification | 20220413 |
Device Code Info | Revolution Apex, Revolution CT with Apex Edition |
Center Classification Date | 20220405 |
Recall Initiation Date | 20210630 |
Recalling Firm | GE Healthcare, LLC |
Initial Notification | Letter |
Similar To |