Beckman Coulter Inc. Recall 90087
Description: MicroScan WalkAway-40 plus Instrument REF B1018-283
Beckman Coulter Inc. Recall 90087 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1083-2022 |
Event ID | 90087 |
Event Description | MicroScan WalkAway-40 plus Instrument REF B1018-283 |
Product Type | Devices |
Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen. |
Quantity | 22 systems |
Recall Reason | WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere |
Device Classification | 20220525 |
Device Code Info | Catalog Number: B1018-283 UDI-DI Code: 1509959065866 Serial Numbers: 34010003 34010004 34010005 34010006 34010007 34010008 34010009 34010010 34010011 34010012 34010013 34010014 34010015 34010016 34010017 34010018 34010019 34010020 34010021 34010022 34010023 34010024 |
Center Classification Date | 20220513 |
Recall Initiation Date | 20191021 |
Recalling Firm | Beckman Coulter Inc. |
Initial Notification | Letter |
Similar To |