Baxter Healthcare Corporation Recall 90322
Description: NaviCare Nurse Call/Voalte Nurse Call
Baxter Healthcare Corporation Recall 90322 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1306-2022 |
Event ID | 90322 |
Event Description | NaviCare Nurse Call/Voalte Nurse Call |
Product Type | Devices |
Distribution | nationwide and Canada |
Quantity | 283 installations |
Recall Reason | An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority. |
Device Classification | 20220706 |
Device Code Info | software versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300 |
Center Classification Date | 20220630 |
Recall Initiation Date | 20220531 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Letter |
Similar To |