K2M, Inc. Recall 90462
Description: Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
K2M, Inc. Recall 90462 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1307-2022 |
Event ID | 90462 |
Event Description | Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments. |
Product Type | Devices |
Distribution | Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to Canada, EU, and Hong Kong. |
Quantity | 53 devices (19 US; 34 OUS) |
Recall Reason | Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue. |
Device Classification | 20220706 |
Device Code Info | UDI-DI: 10888857326804 Lot PCMW |
Center Classification Date | 20220630 |
Recall Initiation Date | 20220613 |
Recalling Firm | K2M, Inc. |
Initial Notification | Letter |
Similar To |