Product Safety Recalls

Product Recall Tracker

Menarini Silicon Biosystems Recall 90417

Description: CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Menarini Silicon Biosystems Recall 90417 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1345-2022
Event ID90417
Event DescriptionCELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001
Product TypeDevices
DistributionCA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,
Quantity484 units
Recall ReasonHigh number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples
Device Classification20220713
Device Code InfoUDI-DI: (01)10888385000016(17)221004(10)S108 Lot Number: S108 Exp. Date: 04Oct2022
Center Classification Date20220707
Recall Initiation Date20220520
Recalling FirmMenarini Silicon Biosystems
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.