Baxter Healthcare Corporation Recall 90848
Description: Simple Clamp, US/EURO Rails, UPPER EXTREMITY DISTRACTION CMPT, REF F-40018, #800101. To position, support and/or distract the patients hand, wrist, forearm, arm, and shoulder in surgical procedures.
Baxter Healthcare Corporation Recall 90848 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0180-2023 |
Event ID | 90848 |
Event Description | Simple Clamp, US/EURO Rails, UPPER EXTREMITY DISTRACTION CMPT, REF F-40018, #800101. To position, support and/or distract the patients hand, wrist, forearm, arm, and shoulder in surgical procedures. |
Product Type | Devices |
Distribution | MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa |
Quantity | 1 unit |
Recall Reason | the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity. |
Device Classification | 20221109 |
Device Code Info | UDI/DI 00615521012731, Lot Number: 22-124 |
Center Classification Date | 20221103 |
Recall Initiation Date | 20220920 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |