LeMaitre Vascular, Inc. Recall 91006

Description: TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by this recall are 3F and 4F..

LeMaitre Vascular, Inc. Recall 91006 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0228-2023
Event ID91006
Event DescriptionTufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by this recall are 3F and 4F..
Product TypeDevices
DistributionNo US distribution. Foreign distribution to United Kingdom, Germany, France, Sweden, Austria, and Italy.
Quantity72 (OUS only)
Recall ReasonThree lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
Device Classification20221123
Device Code InfoLots XSL0097, XSL0098, and XSL0099
Center Classification Date20221111
Recall Initiation Date20220930
Recalling FirmLeMaitre Vascular, Inc.