Datascope Corp. Recall 91652
Description: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Datascope Corp. Recall 91652 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1277-2023 |
Event ID | 91652 |
Event Description | Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85 |
Product Type | Devices |
Distribution | Distribution list not yet provided. |
Quantity | 391 total IABPs |
Recall Reason | There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy. |
Device Classification | 20230329 |
Device Code Info | Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6 |
Center Classification Date | 20230322 |
Recall Initiation Date | 20230228 |
Recalling Firm | Datascope Corp. |
Initial Notification | Letter |
Similar To |