Cytocell Ltd. Recall 92586
Description: RET Distal Probe. Used in Fluorescence in situ hybridization (FISH), a laboratory technique for detecting and locating a specific DNA sequence on a chromosome.
Cytocell Ltd. Recall 92586 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2210-2023 |
Event ID | 92586 |
Event Description | RET Distal Probe. Used in Fluorescence in situ hybridization (FISH), a laboratory technique for detecting and locating a specific DNA sequence on a chromosome. |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of CA, FL, OH, & SC. |
Quantity | 61 units |
Recall Reason | DNAs used during the manufacture of two affected models were accidentally inverted during manufacturing. |
Device Classification | 20230726 |
Device Code Info | Catalog No. LPS509-A; UDI-DI: 05055844900516; Lot No. 076764, 080124, 082129, 082560, 083247, 083550, 083850. |
Center Classification Date | 20230720 |
Recall Initiation Date | 20230524 |
Recalling Firm | Cytocell Ltd. |
Initial Notification | Letter |