Siemens Medical Solutions USA, Inc Recall 92653

Description: ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000

Siemens Medical Solutions USA, Inc Recall 92653 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2361-2023
Event ID92653
Event DescriptionARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Quantity10 units
Recall ReasonDue to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Device Classification20230816
Device Code InfoUDI-DI: 4056869149325 Serial Numbers: Serial 164840 164851 164854 164853 164855 164856 164833 164836 164845 164850
Center Classification Date20230808
Recall Initiation Date20230619
Recalling FirmSiemens Medical Solutions USA, Inc