Siemens Medical Solutions USA, Inc Recall 92653
Description: ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000
Siemens Medical Solutions USA, Inc Recall 92653 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2361-2023 |
Event ID | 92653 |
Event Description | ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom. |
Quantity | 10 units |
Recall Reason | Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system |
Device Classification | 20230816 |
Device Code Info | UDI-DI: 4056869149325 Serial Numbers: Serial 164840 164851 164854 164853 164855 164856 164833 164836 164845 164850 |
Center Classification Date | 20230808 |
Recall Initiation Date | 20230619 |
Recalling Firm | Siemens Medical Solutions USA, Inc |
Initial Notification | Letter |