Siemens Healthcare Diagnostics, Inc. Recall 92800
Description: Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin Test Code: GGT Siemens Material Number (SMN): 11097597
Siemens Healthcare Diagnostics, Inc. Recall 92800 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2473-2023 |
Event ID | 92800 |
Event Description | Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin Test Code: GGT Siemens Material Number (SMN): 11097597 |
Product Type | Devices |
Distribution | US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam |
Quantity | 19544 units |
Recall Reason | Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿ CH HDL Cholesterol (HDLC) |
Device Classification | 20230906 |
Device Code Info | Unique Device Identification (UDI): 00630414596440 All Lot Numbers |
Center Classification Date | 20230828 |
Recall Initiation Date | 20230710 |
Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
Initial Notification | Letter |
Similar To |