Product Safety Recalls

Product Recall Tracker

Ortho-Clinical Diagnostics, INc. Recall 92928

Description: VITROS XT Chemistry Products ALKP Slides

Ortho-Clinical Diagnostics, INc. Recall 92928 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2624-2023
Event ID92928
Event DescriptionVITROS XT Chemistry Products ALKP Slides
Product TypeDevices
DistributionWorldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
Quantity402,966 units
Recall ReasonSamples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
Device Classification20230927
Device Code InfoProduct Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025
Center Classification Date20230921
Recall Initiation Date20230805
Recalling FirmOrtho-Clinical Diagnostics, INc.
Initial Notification Letter