Stryker Corporation Recall 92841

Description: Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.

Stryker Corporation Recall 92841 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2596-2023
Event ID92841
Event DescriptionStryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.
Product TypeDevices
DistributionNationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK
Quantity109 units US; 290 units OUS
Recall ReasonPotential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
Device Classification20230927
Device Code InfoGTIN: 04546540039415 Lot Numbers: 22314012 22336012
Center Classification Date20230918
Recall Initiation Date20230807
Recalling FirmStryker Corporation
Initial Notification Letter