Hamilton Medical, Inc. Recall 92873

Description: Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Hamilton Medical, Inc. Recall 92873 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2529-2023
Event ID92873
Event DescriptionHamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Product TypeDevices
DistributionDistribution US nationwide, including Puerto Rico. OUS: Not provided
Quantity6 units
Recall ReasonDegrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Device Classification20230927
Device Code InfoUDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Center Classification Date20230920
Recall Initiation Date20230726
Recalling FirmHamilton Medical, Inc.