Philips Healthcare Recall 93000
Description: Incisive CT, software version 5.0
Philips Healthcare Recall 93000 Information
Status | Ongoing |
Mandated? | FDA Mandated |
Recall Number | Z-2520-2023 |
Event ID | 93000 |
Event Description | Incisive CT, software version 5.0 |
Product Type | Devices |
Distribution | US Nationwide Distribution |
Quantity | 140 systems |
Recall Reason | Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm. |
Device Classification | 20231011 |
Device Code Info | software version 5.0 |
Center Classification Date | 20230929 |
Recall Initiation Date | 20230802 |
Recalling Firm | Philips Healthcare |
Initial Notification | Letter |
Similar To |