Aizu Olympus Co., Ltd. Recall 93222
Description: Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Models: LF-DP LF-GP LF-TP
Aizu Olympus Co., Ltd. Recall 93222 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0338-2024 |
Event ID | 93222 |
Event Description | Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Models: LF-DP LF-GP LF-TP |
Product Type | Devices |
Distribution | Worldwide - US Nationwide distribution. |
Quantity | 65 units |
Recall Reason | Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection |
Device Classification | 20231129 |
Device Code Info | UDI-DI: LF-DP 04953170136856 LF-GP 04953170061998 LF-TP 04953170317576 |
Center Classification Date | 20231117 |
Recall Initiation Date | 20230925 |
Recalling Firm | Aizu Olympus Co., Ltd. |
Initial Notification | Letter |
Similar To |