Product Safety Recalls

Product Recall Tracker

Maquet Medical Systems USA Recall 93849

Description: BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.

Maquet Medical Systems USA Recall 93849 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1077-2024
Event ID93849
Event DescriptionBEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.
Product TypeDevices
DistributionWorldwide - US Nationwide distribution.
Quantity16520 worldwide; 6657 US
Recall ReasonThe Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
Device Classification20240221
Device Code InfoUDI-DI 04058863080383 Lots 3000280906, 3000280907, 3000280908, 3000280909, 3000282253, 3000282254, 3000282255, 3000291127, 3000300688, 3000314506, 3000316206, 3000316600, 3000317531, 3000317532, 3000318850, 3000318851, 3000318852, 3000318853, 3000319748, 3000319749, 3000319750, 3000319751, 3000319778, 3000319781, 3000323082, 3000323084, 3000323085, 3000324495, 3000324496, 3000324497, 3000324499, 3000325557, 3000325558, 3000325559, 3000327120, 3000327121, 3000328404, 3000328405, 3000334413, 3000334414, 3000336487 ***Added 3/21/24*** 3000330091, 3000330092, 3000330094, 3000330095, 3000330096, 3000333095, 3000333096, 3000333097, 3000334409, 3000334410, 3000336488, 3000337834, 3000342904, 3000342905, 3000342906, 3000342907, 3000342908, 3000344446, 3000344447, 3000344471, 3000344472, 3000344474, 3000346048, 3000346049, 3000346050, 3000346051, 3000346052, 3000348543, 3000348546, 3000348552, 3000348553, 3000349793, 3000349795, 3000349796, 3000349797, 3000352614, 3000352615, 3000352616, 3000352617, 3000353439, 3000353440
Center Classification Date20240213
Recall Initiation Date20240116
Recalling FirmMaquet Medical Systems USA
Initial Notification Letter