Mckesson Medical-Surgical Inc. Corporate Office Recall 94024
Description: Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"
Mckesson Medical-Surgical Inc. Corporate Office Recall 94024 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1313-2024 |
Event ID | 94024 |
Event Description | Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST" |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of AZ, PA and TX. |
Quantity | 25 units |
Recall Reason | Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result. |
Device Classification | 20240327 |
Device Code Info | Catalogue Number: LCC101; UDI/DI: 10724995220072 |
Center Classification Date | 20240319 |
Recall Initiation Date | 20240109 |
Recalling Firm | Mckesson Medical-Surgical Inc. Corporate Office |
Initial Notification | Telephone |
Similar To |