Johnson & Johnson Surgical Vision, Inc. Recall 94231
Description: Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Johnson & Johnson Surgical Vision, Inc. Recall 94231 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1755-2024 |
Event ID | 94231 |
Event Description | Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery. |
Product Type | Devices |
Distribution | U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI. |
Quantity | 3,428 units |
Recall Reason | Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms. |
Device Classification | 20240508 |
Device Code Info | Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923 |
Center Classification Date | 20240502 |
Recall Initiation Date | 20240314 |
Recalling Firm | Johnson & Johnson Surgical Vision, Inc. |
Initial Notification | Letter |
Similar To |