NRT X-RAY A/S Recall 94363
Description: Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
NRT X-RAY A/S Recall 94363 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1784-2024 |
Event ID | 94363 |
Event Description | Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI. |
Quantity | 35 |
Recall Reason | X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator. |
Device Classification | 20240515 |
Device Code Info | REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023 |
Center Classification Date | 20240507 |
Recall Initiation Date | 20240417 |
Recalling Firm | NRT X-RAY A/S |
Initial Notification | Letter |