Philips North America Llc Recall 94508

Description: Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143

Philips North America Llc Recall 94508 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1844-2024
Event ID94508
Event DescriptionEvolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Recall ReasonThe g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Device Classification20240529
Device Code InfoUDI-DI: N/A Serial Numbers: 42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714
Center Classification Date20240517
Recall Initiation Date20240503
Recalling FirmPhilips North America Llc
Initial Notification Letter