Philips North America Llc Recall 94508
Description: Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
Philips North America Llc Recall 94508 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1844-2024 |
Event ID | 94508 |
Event Description | Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom. |
Recall Reason | The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire |
Device Classification | 20240529 |
Device Code Info | UDI-DI: N/A Serial Numbers: 42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714 |
Center Classification Date | 20240517 |
Recall Initiation Date | 20240503 |
Recalling Firm | Philips North America Llc |
Initial Notification | Letter |