MEDLINE INDUSTRIES, LP – Northfield Recall 94537

Description: MEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020

MEDLINE INDUSTRIES, LP – Northfield Recall 94537 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1938-2024
Event ID94537
Event DescriptionMEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
Quantity5,075,234 syringes
Recall ReasonIt has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
Device Classification20240612
Device Code InfoUDI/DI 10080196306183, All lots in the format: 897YYMMXXXX
Center Classification Date20240531
Recall Initiation Date20240422
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield