GE Vingmed Ultrasound As Recall 94772
Description: GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW
GE Vingmed Ultrasound As Recall 94772 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2317-2024 |
Event ID | 94772 |
Event Description | GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW |
Product Type | Devices |
Distribution | US: IL, MA, NM, NY, OK, |
Quantity | 1 unit |
Recall Reason | GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user. |
Device Classification | 20240717 |
Device Code Info | a) H45041SW, UDI/DI *, Serial Numbers: 001423S70 |
Center Classification Date | 20240705 |
Recall Initiation Date | 20240502 |
Recalling Firm | GE Vingmed Ultrasound As |
Initial Notification | Letter |