Product Safety Recalls

Product Recall Tracker

B Braun Medical Inc Recall 94890

Description: Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A

B Braun Medical Inc Recall 94890 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2434-2024
Event ID94890
Event DescriptionBrand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A
Product TypeDevices
DistributionDomestic distribution nationwide. International distribution to Canada.
Quantity9943
Recall ReasonPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Device Classification20240731
Device Code InfoLot Code: Model No 8713031U; UDI-DI 4046964797958; All Serial Numbers
Center Classification Date20240725
Recall Initiation Date20240614
Recalling FirmB Braun Medical Inc
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.