Stryker Corporation Recall 95502
Description: Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001
Stryker Corporation Recall 95502 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0367-2025 |
Event ID | 95502 |
Event Description | Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia. |
Quantity | 2,620 eaches |
Recall Reason | There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries |
Device Classification | 20241120 |
Device Code Info | GTIN: 07613327117219 Lot Number: 64117291 |
Center Classification Date | 20241113 |
Recall Initiation Date | 20241004 |
Recalling Firm | Stryker Corporation |
Initial Notification | Letter |
Similar To |