Product Safety Recalls

Product Recall Tracker

Shimadzu Corp. Analytical & Measuring Instruments Div. Recall 95633

Description: Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Shimadzu Corp. Analytical & Measuring Instruments Div. Recall 95633 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0554-2025
Event ID95633
Event DescriptionAutosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Product TypeDevices
DistributionWorldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
Quantity48
Recall ReasonThere have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Device Classification20241204
Device Code InfoModel/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.
Center Classification Date20241126
Recall Initiation Date20241009
Recalling FirmShimadzu Corp. Analytical & Measuring Instruments Div.